- Table B-4 to Subpart B of Part 53—Line Voltage and Room Temperature Test Conditions
Test day | Line
voltage, 1 rms | Room
temperature, 2 °C | Comments | 0 | 115 | 25 | Initial set-up and adjustments. | 1 | 125 | 20 | 2 | 105 | 20 | 3 | 125 | 30 | Adjustments and/or periodic maintenance permitted at end of tests. | 4 | 105 | 30 | 5 | 125 | 20 | 6 | 105 | 20 | Adjustments and/or periodic maintenance permitted at end of tests. | 7 | 125 | 30 | Examine test results to ascertain if further testing is required. | 8 | 105 | 30 | 9 | 125 | 20 | Adjustments and/or periodic maintenance permitted at end of tests. | 10 | 105 | 20 | 11 | 125 | 30 | 12 | 105 | 30 | Adjustments and/or periodic maintenance permitted at end of tests. | 13 | 125 | 20 | 14 | 105 | 20 | 15 | 125 | 30 |
---|
1 Voltage specified shall be controlled to ±1 volt.
2 Temperatures shall be controlled to ±1 °C.