- Table B-1 to Subpart B of Part 53—Performance Limit Specifications for Automated Methods
Performance
parameter | Units 1 | SO | O | CO | NO (Std. range) | Definitions and test
procedures | Std. range 3 | Lower range | Std. range 3 | Lower range | Std. range 3 | Lower range | 1. Range | ppm | 0-0.5 | <0.5 | 0-0.5 | <0.5 | 0-50 | <50 | 0-0.5 | Sec. 53.23(a) | 2. Noise | ppm | 0.001 | 0.0005 | 0.0025 | 0.001 | 0.2 | 0.1 | 0.005 | Sec. 53.23(b) | 3. Lower detectable limit | ppm | 0.002 | 0.001 | 0.005 | 0.002 | 0.4 | 0.2 | 0.010 | Sec. 53.23(c) | 4. Interference equivalent | Each interferent | ppm | ±0.005 | 4 ±0.005 | ±0.005 | ±0.005 | ±1.0 | ±0.5 | ±0.02 | Sec. 53.23(d) | Total, all interferents | ppm | - | - | - | - | - | - | 0.04 | Sec. 53.23(d) | 5. Zero drift, 12 and 24 hour | ppm | ±0.004 | ±0.002 | ±0.004 | ±0.002 | ±0.5 | ±0.3 | ±0.02 | Sec. 53.23(e) | 6. Span drift, 24 hour | 20% of upper range limit | Percent | - | - | - | - | - | - | ±20.0 | Sec. 53.23(e) | 80% of upper range limit | Percent | ±3.0 | ±3.0 | ±3.0 | ±3.0 | ±2.0 | ±2.0 | ±5.0 | Sec. 53.23(e) | 7. Lag time | Minutes | 2 | 2 | 2 | 2 | 2.0 | 2.0 | 20 | Sec. 53.23(e) | 8. Rise time | Minutes | 2 | 2 | 2 | 2 | 2.0 | 2.0 | 15 | Sec. 53.23(e) | 9. Fall time | Minutes | 2 | 2 | 2 | 2 | 2.0 | 2.0 | 15 | Sec. 53.23(e) | 10. Precision | 20% of upper range limit | - | - | - | - | - | - | Sec. 53.23(e) | Percent 5 | 2 | 2 | 2 | 2 | 1.0 | 1.0 | 4 | Sec. 53.23(e) | 80% of upper range limit | - | - | - | - | - | - | Sec. 53.23(e) | Percent 5 | 2 | 2 | 2 | 2 | 1.0 | 1.0 | 6 | Sec. 53.23(e)
Sec. 53.23(e) |
---|
1 To convert from parts per million (ppm) to µg/m 3 at 25 °C and 760 mm Hg, multiply by M/0.02447, where M is the molecular weight of the gas. Percent means percent of the upper measurement range limit.
2 Tests for interference equivalent and lag time do not need to be repeated for any lower range provided the test for the standard range shows that the lower range specification (if applicable) is met for each of these test parameters.
3 For candidate analyzers having automatic or adaptive time constants or smoothing filters, describe their functional nature, and describe and conduct suitable tests to demonstrate their function aspects and verify that performances for calibration, noise, lag, rise, fall times, and precision are within specifications under all applicable conditions. For candidate analyzers with operator-selectable time constants or smoothing filters, conduct calibration, noise, lag, rise, fall times, and precision tests at the highest and lowest settings that are to be included in the FRM or FEM designation.
4 For nitric oxide interference for the SO
5 Standard deviation expressed as percent of the URL.